The author of the comment rightly suggests the improvement of our data-correction method II by taking into account the dead time of the fast channel of the digital electronics. In our paper, we preferred to neglect the dead time of the fast channel since we did not have a reliable measurement of it, but could only estimate it from the trapezoidal filter length. If we compare the nominal value of the slow channel dead time reported in the XIA manual (about 8.2 ms) and the one we have measured from the data (9.1 ms) we find an evident disagreement; thus, we were not completely confident in using the estimated value of 0.8 ms for the fast channel dead time....

Ciatto, G., D'Acapito, F., Boscherini, F., Mobilio, S. (2004). Comments on treatment of EXAFS data taken in the fluorescence mode in non-linear conditions by G. Ciatto et al. (2004). J. Synchrotron Rad. 11, 278-283. JOURNAL OF SYNCHROTRON RADIATION, 11, 509-509 [10.1107/S0909049504024598].

Comments on treatment of EXAFS data taken in the fluorescence mode in non-linear conditions by G. Ciatto et al. (2004). J. Synchrotron Rad. 11, 278-283

MOBILIO, Settimio
2004-01-01

Abstract

The author of the comment rightly suggests the improvement of our data-correction method II by taking into account the dead time of the fast channel of the digital electronics. In our paper, we preferred to neglect the dead time of the fast channel since we did not have a reliable measurement of it, but could only estimate it from the trapezoidal filter length. If we compare the nominal value of the slow channel dead time reported in the XIA manual (about 8.2 ms) and the one we have measured from the data (9.1 ms) we find an evident disagreement; thus, we were not completely confident in using the estimated value of 0.8 ms for the fast channel dead time....
2004
Ciatto, G., D'Acapito, F., Boscherini, F., Mobilio, S. (2004). Comments on treatment of EXAFS data taken in the fluorescence mode in non-linear conditions by G. Ciatto et al. (2004). J. Synchrotron Rad. 11, 278-283. JOURNAL OF SYNCHROTRON RADIATION, 11, 509-509 [10.1107/S0909049504024598].
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11590/143326
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 4
  • ???jsp.display-item.citation.isi??? 1
social impact