The calibration cost of an instrument is so expensive that is necessary to reduce the number of calibrated points in the scale. There is the necessity to exploit as well as possible the metrological valence of a few calibrated points to exend all over the instrumental range. The choice of the number andthe position of such few points is the problem faced in the work. Utilizing the Shannon sampling procedure in the transposed domain of the input entities of the instrument, with respectto the time, the truncation error of the cardinal series has faced utilizing the Helms and Thomas upper bound evaluation. To enhance the efficiency of the procedure,has been utilized the minimax procedure searching the optimal indication on the better determination. In every case the efficiency of the procedure is lower on the edge of instrumental scale. To upgrade the edge calibration, making as even as possible the calibration, upon the scale,one has been utilazed the generalized sampling theorem by Papoulis that allows to chos the calibration points more frequent atthe edge, with respec of the central part of thescale. An algorithm to guide the calibration procedure is tempted.

Caciotta, M. (2008). Informative Calibration of the Instrumentation. MEASUREMENT, 41(2), 211-218 [10.1016/j.measurement.2006.09.013].

Informative Calibration of the Instrumentation

CACIOTTA, Maurizio
2008-01-01

Abstract

The calibration cost of an instrument is so expensive that is necessary to reduce the number of calibrated points in the scale. There is the necessity to exploit as well as possible the metrological valence of a few calibrated points to exend all over the instrumental range. The choice of the number andthe position of such few points is the problem faced in the work. Utilizing the Shannon sampling procedure in the transposed domain of the input entities of the instrument, with respectto the time, the truncation error of the cardinal series has faced utilizing the Helms and Thomas upper bound evaluation. To enhance the efficiency of the procedure,has been utilized the minimax procedure searching the optimal indication on the better determination. In every case the efficiency of the procedure is lower on the edge of instrumental scale. To upgrade the edge calibration, making as even as possible the calibration, upon the scale,one has been utilazed the generalized sampling theorem by Papoulis that allows to chos the calibration points more frequent atthe edge, with respec of the central part of thescale. An algorithm to guide the calibration procedure is tempted.
2008
Caciotta, M. (2008). Informative Calibration of the Instrumentation. MEASUREMENT, 41(2), 211-218 [10.1016/j.measurement.2006.09.013].
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11590/144185
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 1
  • ???jsp.display-item.citation.isi??? 2
social impact