This paper describes the performance test of the Pulsed Doppler feature of clinical ultrasound platforms of various manufacturers via a blood flow phantom. This is made up by a tissue mimicking material containing a blood mimicking fluid which circulates in a vessel with a constant mean velocity by means of a centrifuge pump. From the digitized images of the Pulsed Doppler spectra over time, the mean and maximum velocities have been computed with a software package developed by the Authors. For laminar flows with a mean velocity ranging from 1,1±0,7 cm/s and 12,7±0,7 cm/s, measurements on the blood mimicking fluid were obtained using 7 different ultrasound scanner platforms, equipped with phased array and convex array probes. The pooled data have shown an overestimation of the mean velocity, from over 200 percent down to about 40 percent, depending on nominal flow rate and ultrasound equipment composition. ©2008 IEEE.

Marinozzi, F., Bini, F., D'Orazio, A., Scorza, A. (2008). Performance tests of sonographic instruments for the measure of flow speed. In IST 2008 - IEEE Workshop on Imaging Systems and Techniques Proceedings (pp.50-55). 345 E 47TH ST, NEW YORK, NY 10017 USA : IEEE [10.1109/IST.2008.4659939].

Performance tests of sonographic instruments for the measure of flow speed

BINI, FABIANO
Membro del Collaboration Group
;
Scorza, Andrea
Writing – Original Draft Preparation
2008-01-01

Abstract

This paper describes the performance test of the Pulsed Doppler feature of clinical ultrasound platforms of various manufacturers via a blood flow phantom. This is made up by a tissue mimicking material containing a blood mimicking fluid which circulates in a vessel with a constant mean velocity by means of a centrifuge pump. From the digitized images of the Pulsed Doppler spectra over time, the mean and maximum velocities have been computed with a software package developed by the Authors. For laminar flows with a mean velocity ranging from 1,1±0,7 cm/s and 12,7±0,7 cm/s, measurements on the blood mimicking fluid were obtained using 7 different ultrasound scanner platforms, equipped with phased array and convex array probes. The pooled data have shown an overestimation of the mean velocity, from over 200 percent down to about 40 percent, depending on nominal flow rate and ultrasound equipment composition. ©2008 IEEE.
2008
9781424424979
Marinozzi, F., Bini, F., D'Orazio, A., Scorza, A. (2008). Performance tests of sonographic instruments for the measure of flow speed. In IST 2008 - IEEE Workshop on Imaging Systems and Techniques Proceedings (pp.50-55). 345 E 47TH ST, NEW YORK, NY 10017 USA : IEEE [10.1109/IST.2008.4659939].
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11590/337549
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 18
  • ???jsp.display-item.citation.isi??? 5
social impact