In this article we present a study about using VNA calibrated or uncalibrated measurements for the purpose of measuring the Q–factor of a dielectric loaded microwave resonator. We evaluate the uncertainties in Q involved in both methods. We use a complete and robust transmission resonant model for the fit and parameter extraction. The Q values obtained from uncalibrated and calibrated data were compared by taking into account the S-parameter measurement uncertainty.
Torokhtii, K., Alimenti, A., Pompeo, N., Leccese, F., Orsini, F., Scorza, A., et al. (2018). Q-factor of microwave resonators: calibrated vs. uncalibrated measurements. In XXII World Congress of the International Measurement Confederation (IMEKO 2018) 3–6 September 2018, Belfast, United Kingdom (pp.052027) [10.1088/1742-6596/1065/5/052027].
Q-factor of microwave resonators: calibrated vs. uncalibrated measurements
K. Torokhtii
;ALIMENTI, ANDREA;N. Pompeo;F. Leccese;F. Orsini;A. Scorza;S. A. Sciuto;E. Silva
2018-01-01
Abstract
In this article we present a study about using VNA calibrated or uncalibrated measurements for the purpose of measuring the Q–factor of a dielectric loaded microwave resonator. We evaluate the uncertainties in Q involved in both methods. We use a complete and robust transmission resonant model for the fit and parameter extraction. The Q values obtained from uncalibrated and calibrated data were compared by taking into account the S-parameter measurement uncertainty.I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.