Quality Controls (QCs) for the evaluation of Color Doppler (CD) performance play a pivotal role since CD Is one of the most widely used Doppler techniques in medical imaging. Among the recommended test parameters, directional accuracy at 90° is included, even if its assessment is carried out visually. Therefore, in this study, a novel CD test parameter, called blind angle, has been proposed and defined as it allows testing the Doppler frequency shift dependency on the insonation angle Moreover, a novel automatic method for the estimation of this parameter through the post-processing of CD videos has been proposed. Data have been collected from a single ultrasound diagnostic system equipped with a phased array probe. Tests have been carried out for two probe frequencies and three constant flow rate regimes set on a flow phantom. Based on the preliminary promising outcomes, further studies are going to be carried out.
Fiori, G., Scorza, A., Schmid, M., Galo, J., Conforto, S., Sciuto, S.A. (2022). A preliminary study on the blind angle estimation for Quality Assessment of Color Doppler ultrasound diagnostic systems. In 25th IMEKO TC-4 International Symposium on Measurement of Electrical Quantities, IMEKO TC-4 2022 and 23rd International Workshop on ADC and DAC Modelling and Testing, IWADC 2022 (pp.325-329). International Measurement Confederation (IMEKO).
A preliminary study on the blind angle estimation for Quality Assessment of Color Doppler ultrasound diagnostic systems
Fiori G.
Investigation
;Scorza A.Project Administration
;Schmid M.;Conforto S.;Sciuto S. A.
2022-01-01
Abstract
Quality Controls (QCs) for the evaluation of Color Doppler (CD) performance play a pivotal role since CD Is one of the most widely used Doppler techniques in medical imaging. Among the recommended test parameters, directional accuracy at 90° is included, even if its assessment is carried out visually. Therefore, in this study, a novel CD test parameter, called blind angle, has been proposed and defined as it allows testing the Doppler frequency shift dependency on the insonation angle Moreover, a novel automatic method for the estimation of this parameter through the post-processing of CD videos has been proposed. Data have been collected from a single ultrasound diagnostic system equipped with a phased array probe. Tests have been carried out for two probe frequencies and three constant flow rate regimes set on a flow phantom. Based on the preliminary promising outcomes, further studies are going to be carried out.I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.